Development and qualification of a VH-TDMA for the study of pure aerosols

Show simple item record

dc.contributor.author Oxford, Christopher R.
dc.contributor.author Rapp, Charles M.
dc.contributor.author Wang, Yang
dc.contributor.author Kumar, Purushottam
dc.contributor.author Watson, Daniel
dc.contributor.author Portelli, Julianna L.
dc.contributor.author Sussman, Eric A.
dc.contributor.author Dhawan, Steven
dc.contributor.author Jiang, Jingkun
dc.contributor.author Williams, Brent J
dc.date.accessioned 2019-06-18T10:49:52Z
dc.date.available 2019-06-18T10:49:52Z
dc.date.issued 2018-11
dc.identifier.citation Oxford, Christopher R.; Rapp, Charles M.; Wang, Yang; Kumar, Purushottam; Watson, Daniel; Portelli, Julianna L.; Sussman, Eric A.; Dhawan, Steven; Jiang, Jingkun and Williams, Brent J., "Development and qualification of a VH-TDMA for the study of pure aerosols�, Aerosol Science and Technology, DOI: .1080/02786826.2018.1547358, Nov. 2018. en_US
dc.identifier.issn 0278-6826
dc.identifier.issn 1521-7388
dc.identifier.uri https://doi.org/10.1080/02786826.2018.1547358
dc.identifier.uri https://repository.iitgn.ac.in/handle/123456789/4436
dc.description.abstract We create and qualify a Volatility and Hygroscopicity Tandem Differential Mobility Analyzer (VH-TDMA) for the study of aerosols. This VH-TDMA measures size distributions, volatility, and hygroscopicity and includes an auxiliary conditioner that allows quick connection to other external aerosol conditioners. The differential mobility analyzers are not temperature controlled, allowing the surrounding environment to influence the measurement conditions, and this is fully accounted for when measuring aerosol volatility and hygroscopicity. For the volatility conditioner, the VH-TDMA uses a 15?m coil of tubing in an oven to evaporate aerosol samples at elevated temperatures. We measured several single component model aerosols to qualify the differential mobility particle sizer (DMPS) channel and each of the conditioners: hygroscopicity and volatility. Due to insufficient power supply calibration in this study, the TDMA channel is limited to particle sizes greater than 70?nm. The DMPS channel was able to reproduce ammonium sulfate size distributions when compared to common scanning mobility particle sizers. For hygroscopicity, the standard deviation in the measured ammonium sulfate growth factors was 0.03 over a 4-h experiment. From this data, the TDMA has an observed relative humidity error of �0.6% with manufacturer reported error of �1.2% relative humidity. The volatility channel reproduced the previously published saw tooth pattern of room temperature saturation vapor pressures from atomized C3-C9 diacids. The maximum percent difference in room temperature saturation vapor pressure was approximately 80%. The enthalpy of sublimation derived from the diacids increased monotonically (except for suberic acid) and resembled measurements from mass effusion techniques.
dc.description.statementofresponsibility by Christopher R. Oxford, Charles M. Rapp, Yang Wang , Purushottam Kumar, Daniel Watson, Julianna L. Portelli, Eric A. Sussman ORCID Icon, Steven Dhawan, Jingkun Jiang and Brent J. Williams
dc.language.iso en en_US
dc.publisher Taylor & Francis en_US
dc.title Development and qualification of a VH-TDMA for the study of pure aerosols en_US
dc.type Article en_US
dc.relation.journal Aerosol Science and Technology